Abstract

Techniques for the assessment of internally incorporated radionuclides are essential for monitoring potentially exposed workers to radiation as well as for the preparedness of response in the case of Radiological Emergencies. The Whole Body Counter at Instituto Tecnológico e Nuclear (ITN), Portugal, provides a direct way of measuring the activity of internally incorporated radionuclides in possibly contaminated persons. This equipment is calibrated in energy and efficiency, with a calibration source, and is able to detect and quantify several radionuclides with gamma emissions between 60 and 1900 keV. Also, the Minimum Detectable Activity values for several nuclides of interest were determined, since these values are an indication of the sensitivity of the detection system. Finally, in order to assure reliable measurements, a quality assurance routine has been implemented, consisting of routinely monitoring several parameters of chosen radionuclides with gamma emission energies lying in the energy range of interest, and also for the background. The equipment is now fully operational and a set of routine measurements for the monitoring of exposed personnel, working in different areas of activity, is currently being prepared.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.