The Defence Science and Technology Laboratory (Dstl) has type-tested the Harshaw DXT-RAD dosemeter against the International Standard ISO 12794:2000(E) ‘Individual thermoluminescence dosemeters for extremities and eyes’ [International Organization for Standardization, 2000. International Standard ISO 12794:2000(E). Nuclear energy—radiation protection—individual thermoluminescence dosemeters for extremities and eyes]. The tests carried out included batch homogeneity, reproducibility, linearity, detection threshold, environmental stability (temperature, humidity and light), self irradiation, isotropy, energy response, dose–rate response and neutron response. The DXT-RAD dosemeter satisfied all the requirements of ISO 12794:2000(E), with the exception of its response at low photon energies and intermediate beta energies. However, it should be noted that other forms of the DXT-RAD dosemeter with a thinner entrance window are available to give an improved response to low-energy radiations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.