Abstract

Monitoring and protecting of occupational eye doses in interventional radiology (IR) are very important matters. DOSIRIS™ is the useful solution to estimate the 3 mm dose-equivalent (Hp(3)), and it can be worn behind lead glasses. And DOSIRIS™, adjustable according to 3 axes, it is ideally placed as close to the eye and in contact with the skin. So, DOSIRIS™ will be suitable eye lens dosimeter. However, the fundamental characteristics of the DOSIRIS™ in the diagnostic x-ray energy domain (including that of IR x-ray systems) remain unclear. Here, we evaluated the performance of the dosimeter in that energy range. As a result, the DOSIRIS™ has good fundamental characteristics (batch uniformity, dose linearity, energy dependence, and angular dependence) in the diagnostic x-ray energy domain. We conclude that the DOSIRIS™ has satisfactory basic performance for occupational eye dosimetry in diagnostic x-ray energy settings (including IR x-ray systems).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call