Abstract

A potentially implantable glucose biosensor for continuous monitoring of glucose levels in diabetic patients has been developed. The glucose biosensor is based on an amperometric oxygen electrode and glucose oxidase immobilized on carbon powder held in the form of a liquid suspension. The enzyme material can be replaced (the sensor recharged) without sensor disassembly. Diffusion membranes made of silastic coatings over a microporous polycarbonate membrane are used. Calibration curves of the sensors in phosphate buffer solution and in undiluted blood plasma at body temperature have been obtained. The reproducibility of the sensor response in serum at body temperature is demonstrated. The sensors have a stable signal during storage and continuous operation at body temperature for a period of at least one month.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.