Abstract

A potentially implantable glucose biosensor for continuous monitoring of glucose levels in diabetic patients has been developed. The glucose biosensor is based on an amperometric oxygen electrode and glucose oxidase immobilized on carbon powder held in a form of a liquid suspension. The enzyme material can be replaced (the sensor recharged) without sensor disassembly. Recharging of the biosensor is achieved by injecting fresh immobilized enzyme into the sensor using a septum. Diffusion membranes made of silastic latex-rubber coatings over a microporous polycarbonate membrane are used. Calibration curves of the amperometric signal show linearity over a wide range of glucose concentrations-up to 500 mg/dL (28 mM), covering hypoglycemic, normoglycemic, and hyperglycemic conditions. Preliminary in vitro studies of the biosensor show stable performance during several recharge cycles (of 14 days each) over a period of 4 months. (c) 1994 John Wiley & Sons, Inc.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.