Abstract

High-energy photon source (HEPS) is a 6 GeV ultralow emittance storage ring light source to be built in Beijing, China. Both the horizontal and vertical beam sizes of the HEPS storage ring are below 10 µm. It is a challenge to measure such a small beam size in both directions. To this end, measurement by a Kirkpatrick–Baez (KB) mirror imaging system was evaluated. A test KB system for the Shanghai Synchrotron Radiation Facility storage ring was designed and tested. Two crossed cylindrical mirrors were used to image the dipole source point. Both mirrors can be moved in and out so that the monitor is interchangeable with the original X-ray pinhole monitor. The aberration and point spread function, which would cause image blur, were evaluated. A beam-based calibration scheme was used by varying the beam size with different quadrupole settings and fitting them with the corresponding theoretical values. We updated the original X-ray camera with a new camera having a 5-μm-thick LuAG/Ce scintillator, and the imaging result shows greatly decreased image blur.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.