Abstract

Hemoglobin is a very important physiological index in human blood, especially for the evaluation of the physical performance of athletes. This paper reports an amperometric biosensor for the determination of blood hemoglobin, using potassium ferricyanide as a mediator, based on thin-film technology in a two-electrode system. The biosensor can be used for the determination of hemoglobin at concentrations in a linear range of 10-3000 microM, and the correlation coefficient is 0.996, the relative standard deviation is 2.3% for six determinations for 2000 microM hemoglobin concentration. And the linear range of hemoglobin is studied using methylene blue as a mediator.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.