Abstract

To establish an inductively coupled plasma mass spectrometry(ICP-MS) method for determination of 30 trace elements including As, Ba, Be, Bi, Ni, Cd, Co, Cr, Cs, Cu, Ga, Mn, Pb, Sr, Tl, V, Ge, Mo, Nb, Ti, W, Te, Se, Zr, In, Sb, Hg, Ce, La, and Sm in human blood. The blood samples were analyzed by ICP-MS after diluted 1/10 with 0.01% Triton-X-100 and 0.5% nitric acid solution. Y, Rh and Lu were selected as internal standard in order to correct the matrix interference of Cr, As, Se, and Hg by a hex pole-based collision-reaction cell. Other elements were determined with standard method. The limits of detection, precision and accuracy of the method were evaluated. The accuracy was validated by the determination of the whole blood reference material. All the 30 trace elements have good linearity in their determination range, with the correlation coefficient > 0.9999. The limits of detection of the 30 trace elements were in the range of 1.19 - 2.15 µg/L and the intra-precision and inter-precision (relative standard deviation, RSD) were less than 14.3% (except Hg RSD < 21.2%, and Ni RSD < 15.4%). The spiked recovery for all elements fell within 59.3% - 119.2%. Among the 13 whole blood reference materials, V, Cr, Mn, Co, Ni, Cu, As, Se, Cd, Te, and Pb (1.45, 1.19, 18.40, 0.18, 1.57, 591.00, 2.97, 61.00, 0.35, 1.86, and 9.70 µg/L respectively) fell within the acceptable range and the detection results of Hg (0.59 µg/L) and Mo (1.59 µg/L) were slightly beyond the range. This method was simple, fast and effective. It can be used to monitor the multi-elementary concentration in human blood.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call