Abstract

To establish a method of determining more elements in whole blood by inductively coupled plasma mass spectrometry (ICP-MS). The whole blood samples were treated by 5% nitric acid to remove the proteins, then were centrifuged. Sixteen elements (Be, Al, Mn, V, Cr, Co, Ni, As, Mo, Ag, Cd, Sb, Ba,Tl, Pb and U) in the supernatant were directly measured by ICP-MS. The detection limits of 16 elements were 0.01 ∼ 6.51 µg/L. The linear correlation coefficient was ≥ 0.999. The relative standard deviations were below 5 %. The recovery rates were 105% ∼ 115%. Seronorm Trace Elements SerumL-1 LOT 0903106 and GBW09139g or GBW09140g were used in the quality control, the detected results corresponded with the standard values. ICP-MS technique is a simple, rapid, accurate and reliable method, which can be used to measure several trace elements in whole blood samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.