Abstract

Liquid Xe as a scintillator has a very fast response. The scintillation light decays with 2.2 ns and 22 ns, and the recombination light for gamma-rays decays with 45 ns. Also, the scintillation yield is very high, comparable to that of NaI(Tl) crystals. These properties are considered to be advantageous for application to time of flight-positron emission tomography (TOF-PET). From this perspective, a prototype for a TOF-PET system which consists of two identical liquid xenon detectors has been constructed using liquid Xe as scintillator material and its energy resolution, position resolution and the time resolution were measured. The energy resolutions of 29.9% and 32.4% and position resolutions of 4.8 mm and 9.4 mm were obtained for the two identical detectors, respectively, and a time resolution of 650 ps between the two detectors was obtained for 511 keV gamma rays from positron annihilation. The future development of liquid Xe TOF-PET is discussed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.