Abstract

The performance of a photon detector consisting of two lead glass arrays and three planes of proportional tubes is reported. The detector was designed to measure the energies and positions of photons in the energy range 3–50 GeV. An energy resolution of 1.0% + 5.0% √E (standard deviation) and a spatial resolution approaching 3 mm (standard deviation) were achieved during calibration with a low intensity positron beam. The energy resolution was degraded during use with a high intensity hadron beam whereas the spatial resolution was only slightly affected.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.