Abstract

Objective To compare the performance of the abdominal handheld point-of-care ultrasonography (POCUS) Butterfly-iQ to gold standard transvaginal ultrasonography (US) in identifying the position of intrauterine devices (IUDs) in the hands of a medical doctor specialised in ultrasonography. Methods In this diagnostic accuracy study, a single operator conducted abdominal POCUS followed by conventional transvaginal US. Seventy patients utilising copper or hormonal IUDs were assessed between June 2021 and October 2022. IUDs were categorised as entirely within the uterine cavity or malpositioned. Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and accuracy were calculated for detecting malpositioned IUDs, with conventional US results serving as the reference standard. Concordance rate and Kappa coefficient were computed to assess the agreement between the two ultrasound modalities. Results Among the 70 patients, 46 (65.7%) used copper IUDs, and 24 (34.3%) used hormonal IUDs. Conventional transvaginal US showed IUDs entirely within the uterine cavity in 56 (80%) patients and 14 (20%) IUDs were malpositioned. Of the 14 malpositioned IUDs seen by conventional US, POCUS identified 13 demonstrating a sensitivity of 92.9% (66.1–99.8). Of the 56 IUDs entirely within the uterine cavity shown by conventional US, only two cases were considered malpositioned by POCUS demonstrating a specificity of 96.4% (87.7–99.6). The concordance rate was 95.7%, and the Kappa value was 0.87 in differentiating between IUDs entirely within the uterine cavity and those that were malpositioned. Conclusion Abdominal POCUS using Butterfly-iQ, when administered by an imaging specialist, exhibited excellent performance in confirming IUDs entirely within the uterine cavity.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call