Abstract
The purpose is to compare full-field digital mammography (FFDM), digital breast tomosynthesis (DBT) and contrast-enhanced spectral mammography (CESM) technologies on three mammography systems in terms of image quality and patient dose. Two Senographe Essential with DBT and CESM (denoted S1 and S2) and one Selenia Dimensions (S3) with FFDM and DBT were considered. Dosimetry methods recommended in the European protocol were used. Image quality was tested with CDMAM in FFDM and DBT and with ideal observer method in FFDM. Mean values of mean glandular dose (MGD) from whole patient samples on S1, S2 and S3 were as follows: FFDM 1.65, 1.84 and 2.23mGy; DBT 2.03, 1.96 and 2.87mGy; CESM 2.65 and 3.16mGy, respectively. S3 exhibited better low-contrast detectability for the smallest sized discs of CDMAM and ideal observer in FFDM, and for the largest sized discs in DBT, at similar dose levels.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have