A chemical imaging time-of-flight secondary ion mass spectrometer is described. It consists of a liquid metal ion gun, medium energy resolution reflectron mass analyzer, liquid nitrogen cooled sample stage, preparation chamber and dual stage entry port. Unique features include compatibility with laser postionization experiments, large field of view, cryogenic sample handling capability and high incident ion beam current. Instrument performance is illustrated by the characterization of scanning electron microscopy grids, silver and functionalized polystyrene beads and the postionization of an organic overlayer on a gold substrate.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.