Abstract
Introduction The usefulness of anti-scatter grid in digital pediatric chest radiography was investigated by a phantom study assessing image quality using subjective evaluations and dose measurements. Method A contrast detail study was undertaken. Images were taken with and without anti-scatter grids (grid ratio 15:1), for PMMA phantom thickness ranging from 1 to 15 cm. Fourteen radiographers, experienced in digital pediatric and/or skeletal radiography, read 96 images. At each thickness, the image with and without grid technique was read three times, and in random order. Correct Observation Ratio (COR) and Image Quality Figures (IQF) were calculated. Dose Area Products were measured with DAP-meter. Results The COR and IQF values indicated statistically significant ( p < 0.05) enhanced image quality when using anti-scatter grid for all thicknesses 5 cm and more. The DAP values increased significantly by grid techniques ( p = 0.01). Conclusions Grid techniques may be beneficial to image quality for PMMA objects over 5 cm, but this leads to a higher dose. Although the magnitude of dose may be low, the dose should be kept at a minimum. In conclusion, improving image quality should be based on optimizing post-processing settings and image evaluation conditions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.