Abstract

The present work aimed to optimize exposure settings in pediatric digital chest radiography (DR) with regard to image quality and radiation dosage. A pediatric phantom was imaged with a portable DR unit to examine different exposure settings (range: 75-109 kVp; 0.3-1.28 mAs) for patients of 10-20 kg. Fourteen experienced radiologists compared the structural image quality of the images with a reference image (85 kVp/1.28 mAs). A multiple-reader multiple-case analysis of the radiologists' interpretations was performed. Effective dose was computed and standardized exposure indices (EIs) were extracted for the different exposure settings. For the lowest tube voltage setting (75 kVp/1.28 mAs), radiation dosage could be reduced with 35% relative the reference settings without compromising image quality (p > 0.05). EI was within recommendations (250 ± 100). Lower tube voltage in pediatric DR permitted a dose reduction at maintained quality for the physical aspects and diagnostic performance. Other weight-classes should be examined and adjusted accordingly.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.