Abstract
In CIEF analysis, sample peaks can be identified by their relative peak positions (RPP) that are determined using only two internal pI markers. The two internal pI marker peaks should bracket, as close as possible, the sample peaks. The RPP values of the sample peaks are then calculated using the pI values, peak positions of the two pI markers, and peak position of the sample. Use of this method can effectively compensate for pH gradient distortions that often occur as a result of salts. Also, as shown by the results of this paper, regardless of the linearity of the pH gradient established by the given carrier ampholytes, sample peaks can be identified within an SD of 0.1 pH unit in RPP (<2% RSD) as long as the sample is run using the same carrier ampholytes and maintaining salt concentrations in the range of 0-15 mM.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.