Abstract

Serum chrome and cobalt was measured in 50 patients with trapeziometacarpal total joint replacement with metal-on-metal articulation and compared with serum chrome and cobalt values in 23 patients with trapeziometacarpal total joint replacement with metal-on-polyethylene articulation. In 10 of 50 (20%) patients with metal-on-metal articulation, slightly elevated serum chrome or cobalt values were found compared with only one in 23 (4%) patients with metal-on-polyethylene articulation. All metal values were lower than accepted 'normal values' for metal-on-metal hip arthroplasty and so considered not to be a general health risk. However, the mean disabilities of the arm, shoulder and hand (DASH) score was 24 in patients with elevated serum chrome or cobalt compared with 10 in patients with normal metal values (p < 0.05) suggesting a local clinical effect of the elevated serum chrome or cobalt values. We recommend that patients with trapeziometacarpal total joint replacement with metal-on-metal articulation are followed with DASH score and radiological examination every 3-5 years and serum chrome and cobalt should be analysed in symptomatic cases to learn more about possible local complications leading to, or arising from, metal debris.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.