Abstract
Implantable cardioverter-defibrillators (ICDs) have revolutionized the prognosis for patients at elevated risk of ventricular tachyarrhythmias. For safety, defibrillation should be effective with a minimum of 10 J below the device's maximum energy. While modern ICDs rarely deliver ineffective shocks in primary prevention, the surge in managing severe heart failure patients has led to an increased number of patients with high defibrillation thresholds (DFTs). This article elucidates the potential causes of high DFT, including clinical factors, lead and device placement, the presence of a Left Ventricular Assist Device (LVAD), prolonged ventricular arrhythmias, shock vectors, waveform tilt, medications, and manufacturer-specific options. We also detail management strategies, highlighting alternative shock coil placements, practical recommendations, and case studies from our institution. Our management algorithm suggests addressing preventable causes, re-evaluating coil positions, considering non-invasive system modifications, upgrading to a higher-capacity device, and adding extra coil(s).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.