Organ and effective doses to 90 patients undergoing some selected fluoroscopic examinations at the Korle-Bu Teaching Hospital were estimated using the Monte Carlo-based program (PCXMC version 1.5). Radiation dose was estimated from free-in-air measurements. The mean effective doses were found to be 0.29 +/- 0.07, 0.84 +/- 0.13, 3.15 +/- 0.44, 6.24 +/- 0.70 and 0.38 +/- 0.05 mSv for urethrogram, barium swallow, barium meal, barium enema and myelogram examinations, respectively. The dose area product was estimated to be 3.55 +/- 0.95, 16.44 +/- 2.60, 50.81 +/- 7.04, 99.69 +/- 10.85 and 9.32 +/- 0.99 Gy cm(2) for urethrogram, barium swallow, barium meal, barium enema and myelogram examinations, respectively. Optimisation of procedures is required for barium enema and barium meal examinations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.