Abstract

Prosthetic heart valves have been responsible for extending the life spans and improving the quality of life of many people with serious heart conditions. Even though the heart valves are extremely reliable, eventually they are susceptible to the long-term fatigue and structural failure effects expected for mechanical devices operating over long periods of time. In [2] a classification procedure was developed using spectral features obtained from acoustic signals to determine the condition of the prosthetic heart valve. Although this classification procedure has produced very encouraging results, this method still lacks a fundamental physical description of the sounds produced by the valve during normal operation. In order to obtain a better understanding of the valve acoustic response, we have performed a set of anechoic tests. In this paper, we describe the anechoic experiment and also present limited transient response results. This transient information will eventually be used to identify and improve the features used to classify the valve condition.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.