Abstract
BackgroundThis study presents a novel technique to develop an equivalent circuit model (ECM) for analyzing the responses of the layered body structure to transcutaneous electrical nerve stimulation (TENS) by parameterizing electrical and geometrical properties.Many classical ECMs are non-parametric because of the difficulty in projecting intrapersonal variability in the physical properties into ECM. However, not considering the intrapersonal variability hampers patient-specifically analyzing the body response to TENS and personal optimization of TENS parameter design. To overcome this limitation, we propose a tissue property-based (TPB) approach for the direct parameterization of the physical properties in the layered body structure and thus enable to quantify the effects of intrapersonal variability.ResultsThe proposed method was first validated through in vitro phantom studies and then was applied in-vivo to analyze the TENS on the forearm. The TPB-ECM calculated the impedance network in the forearm and corresponding responses to TENS. In addition, the modelled impedance was in good agreement with well-known impedance properties that have been achieved empirically.ConclusionsThe TPB approach uses the parameterized circuit components compared to non-parametric conventional ECMs, thus overcoming the intrapersonal variability problem of the conventional ECMs. Therefore, the TPB-ECM has a potential for widely-applicable TENS analysis and could provide impactful guidance in the TENS parameter design.
Highlights
This study presents a novel technique to develop an equivalent circuit model (ECM) for analyzing the responses of the layered body structure to transcutaneous electrical nerve stimulation (TENS) by parameterizing electrical and geometrical properties.Many classical ECMs are non-parametric because of the difficulty in projecting intrapersonal variability in the physical properties into ECM
ECM developed with tissue property-based (TPB) approach The 2D-to-3D expansion of partial capacitance (PC) method showed its reliability by verifying the geometric factor (GF) terms are dominantly affected by tissue layer thickness, the geometrical tissue property (Fig. S2)
The coefficients of the regression models of partial capacitance (PPC) and serial partial capacitance (SPC) Interdimensional factor (IF) were calculated by fitting to the 3D simulation data (Fig. 1)
Summary
This study presents a novel technique to develop an equivalent circuit model (ECM) for analyzing the responses of the layered body structure to transcutaneous electrical nerve stimulation (TENS) by parameterizing electrical and geometrical properties.Many classical ECMs are non-parametric because of the difficulty in projecting intrapersonal variability in the physical properties into ECM. Not considering the intrapersonal variability hampers patient- analyzing the body response to TENS and personal optimization of TENS parameter design. To overcome this limitation, we propose a tissue property-based (TPB) approach for the direct parameterization of the physical properties in the layered body structure and enable to quantify the effects of intrapersonal variability. Advances to the ECM are needed for analyzing the body responses by discovering the straightforward relationship between the body impedance and the physical tissue properties
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have