Abstract
A novel palpation-based tumor detection method for robot-assisted minimally invasive surgery (RMIS) is proposed in this paper. A tactile stiffness sensor based on piezoelectric vibration is designed to detect tissue stiffness just by gentle contacting, without the risk of injuring the organ. A novel multi-tumor detection method is proposed to solve the problem that existing tumor detection methods can only detect a single tumor. In addition, the moving distance of the sensor during palpation is introduced into the sampling strategy function as a new factor, so that the total moving distance of sensor can be considered and optimized during the tumor palpation process. Simulation studies are carried out to compare the new method with existing methods in tumor detection. It shows that the proposed method can identify and locate multiple tumors successfully with the highest performance (F1 score > 0.99), and the total moving distance during palpation is reduced by 43% without any compromise in the detection accuracy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.