Abstract

Purpose The aim of this study was to calculate the dosimetric parameters derived from percent depth dose curves by using RPL-GD and TLD-100 detectors for high-energy electron beams. Methods In this study, Monte Carlo simulation code MCNP5 was using to calculate the percent depth dose (PDD) in water phantom for 9, 12, 16 and 20 electron beam energies with utilizing the Radiophotoluminescent (FD-7glass series) and Thermoluminescent (TLD-100) materials. Results We found a good agreement with literature for Monte Carlo calculations parameters derived from the PDD curve such as various ranges R 100 , R 90 , R 80 , R 50 and R p and energies specification such us most probable energy E p , o and mean incident energy E 0 . Conclusion In conclusion, RPL-GD and TLD-100 dosimeters are a very useful detector for routine QA checks for electron therapy [1] .

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.