Abstract

We have investigated the performance of a new gas exchange device (GED), named L001, specifically devised for the ozonation of human blood during extracorporeal circulation. This procedure, defined with the acronym "EBOO," means "extracorporeal blood oxygenation-ozonation." The innovative GED is made of microporous, ozone-resistant, polipropylene hollow fibers with an external diameter of 200 microm, a thickness of 50 microm, and a membrane surface area of 0.22 m(2). The material is coated with phosphorylcholine on the external side in contact with the circulating blood, while a gas mixture, necessarily composed of medical oxygen and ozone (about 99 and 1%, respectively), flows inside the fibers in opposite direction. The new GED has been tested by using a buffered saline solution containing KI and by varying several parameters, and it has shown to be very versatile and efficient. Its main characteristics are minimal foreign surface contact, high gas transfer, and negligible priming volume. This device appears to be a practical, nontoxic, and rather inexpensive tool for performing ozonation of blood for already defined human diseases.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.