Abstract

Venoplasty allows the addition or replacement of leads despite subtotal or total subclavian occlusion. The threshold of the LV pacing lead implanted for biventricular pacing over a period of 18 months increased to greater than 5 V. A pre implant venogram revealed total subclavian occlusion. Venous access was maintained by extraction of the 4 F LV lead over a wire. Subsequently the sheath would not advance despite 6mm balloon inflation to 30 atm with no residual waist. A wire was placed beside the balloon and the balloon was reinflated. The subclavian obstruction was eliminated without damage to the existing leads. The obstruction formed by the fibrous track around an extracted lead may persist despite what appears to be successful balloon dilation. Inflation with a wire beside the balloon increases the effect eliminating the resistant obstruction without damaging the leads.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.