Abstract
A procedure for determining of the isotope ratio 235U/238U in UF6 samples was established using a quadrupole mass spectrometer with ionization by electron impact. The following items were optimized in the spectrometer: the parameters in the ion source that provided the most intense peak, with good shape, for the most abundant isotope; the resolution that reduced the non linear effects and the number of analytical cycles that reduced the uncertainty in the results. The measurement process was characterized with respect to the effects of mass discrimination, linearity and memory effect.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.