Abstract

The coupling of a Rydberg electron capture ion source with a Nermag R10-10H quadrupole mass filter is described. Details are given of the addition to this instrument of a creation cell for atoms excited in Rydberg states. Within the Nermag ion source, such atoms allow attachment of electrons of well-defined thermal energy. SF(6) was used for optimization of the main experimental parameters (gas pressures and voltages applied to the electrodes). The procedure by which Rydberg electron attachment was confirmed is described. A polychlorobiphenyl compound was used to illustrate the performance of this ionization technique. Ion formation was observed in the absence of fragmentation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.