Optical bone densitometry (OBD) has been developed for the early detection of osteoporosis. In recent years, machine learning (ML) techniques have been actively implemented for the areas of medical diagnosis and screening with the goal of improving diagnostic accuracy. The purpose of this study was to verify the feasibility of using the combination of OBD and ML techniques as a screening tool for osteoporosis. Dual energy X-ray absorptiometry (DXA) and OBD measurements were performed on 203 Thai subjects. From the OBD measurements and readily available demographic data, machine learning techniques were used to predict the T-score measured by the DXA. The T-score predicted using the Ridge regressor had a correlation of r = 0.512 with respect to the reference value. The predicted T-score also showed an AUC of 0.853 for discriminating individuals with osteoporosis. The results obtained suggest that the developed model is reliable enough to be used for screening for osteoporosis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.