Abstract

To determine the normal pattern of cervical spinal cord motion with measurement of cervical spinal cord velocity by means of phase-contrast magnetic resonance (MR) imaging. Spinal cord velocity was measured in 11 healthy subjects with a modified gradient-echo pulse sequence on a conventional 1.5-T MR imaging system that generated phase images sensitive to slow motion. Prospective electrocardiogram gating was used to assess velocity as a function of the cardiac cycle. The accuracy of velocity measurements was estimated with images of a phantom moving at constant velocity. The cervical spinal cord moves with an oscillatory pattern in the craniocaudal direction. The maximum velocity (7.0 mm/sec +/- 1.4 [standard deviation]) in the caudal direction occurred approximately 109 msec +/- 20 after electrical cardiac systole. The maximum velocities in subsequent oscillations decreased toward zero before the next cardiac systole. The cervical spinal cord oscillates in a craniocaudal direction after each cardiac systole.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.