Abstract

A signal conditioning circuit based on a relaxation oscillator is proposed for use with resistive bridge sensors. The circuit provides a rectangular-wave output whose frequency is related to the bridge unbalance, and duty-cycle is a function of the overall sensor bridge resistance, hence of the sensor operating temperature. In this way, two measurement values are simultaneously and independently carried on the same output signal. The circuit makes use of a constant current bridge excitation which enables the connection of remote sensors without accuracy degradation, and moreover, for silicon piezoresistive sensors, provides a first-order temperature compensation. A frequency-doubling output stage significantly reduces the nonlinearity due to switching delay times, at the parity of output center frequency and span. Experimental results are reported on the characterization of both the circuit alone and interfaced to a silicon pressure sensor.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.