Abstract

Abstract Organic scintillators with long luminescent lifetimes can theoretically be used to temporally filter out radiation-induced luminescence and Cerenkov light (the so-called stem signal) when used as fibre-coupled radiotherapy dosimeters. Since the medical linear accelerators (linacs 1 ) used for radiotherapy treatments deliver pulsed beams, the stem signal can be suppressed using dosimeter materials with luminescent lifetimes much longer than that of the stem signal. However, producing organic scintillators with long luminescent lifetimes has proven difficult in practice. We report on the results of experiments performed using two organic scintillators, one commercially available and one custom made. The luminescent lifetimes of the scintillators have been measured using i) optical excitation by pulsed UV light, and ii) irradiative excitation using high-energy X-rays from a linac. A luminescent lifetime component on the order of 20 μs was estimated for the custom-made organic scintillator, while the commercial scintillator exhibited a fast component of approximately 5 ns lifetime (7 ns as stated by the manufacturer) and an approximate 10 μs lifetime slow component. Although these lifetimes are not long enough for practical applications in radiotherapy dosimetry, this study supports that the stem signal can be greatly reduced by applying a temporal gating.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.