Abstract

Organ-mounted robots adhere to the surface of a mobile organ as a platform for minimally invasive interventions, providing passive compensation of physiological motion. This approach is beneficial during surgery on the beating heart. Accurate localization in such applications requires accounting for the heartbeat and respiratory motion. Previous work has described methods for modeling quasi-periodic motion of a point and registering to a static preoperative map. The existing techniques, while accurate, require several respiratory cycles to converge. This paper presents a general localization technique for this application, involving function approximation using radial basis function (RBF) interpolation. In an experiment in the porcine model in vivo, the technique yields mean localization accuracy of 1.25mm with a 95% confidence interval of 0.22mm. The RBF approximation provides accurate estimates of robot location instantaneously.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.