Abstract

The authors measured organ radiation doses during multi-slice computed tomography (MSCT) chest scans using a China Sichuan anthropomorphic phantom (CDP-1C). Chest CT images from live volunteers based on automatic tube current modulation (ATCM) techniques were similar to those obtained using the CDP-1C phantom, indicating that the phantom accurately modelled the anatomic structure and X-ray absorbance of the human torso. Indeed, attenuation values differed by <5%. Organ radiation doses were measured using thermoluminescence dosemeters in the CDP-1C. With increased noise index, the CT dose index, the dose-length product and the average organ dose all decreased. Thus, the CDP-1C phantom can also assess dose levels during CT examinations in Chinese patients. The noise index (based on ATCM techniques) should be set to 8.5 or higher to reduce X-ray exposure while maintaining appropriate resolution for diagnosis.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call