Abstract

We have carried out an optoelectronic measurement of the impulse response of an ultrafast sampling oscilloscope with a nominal bandwidth of 100 GHz within a time window of approximately 100 ps. Our experimental technique also considers frequency components above the cut-off frequency of higher order modes of the 1.0 mm coaxial line, which is shown to be important for the specification of the impulse response of ultrafast sampling oscilloscopes. Additionally, we have measured the reflection coefficient of the sampling head induced by the mismatch of the sampling circuit and the coaxial connector which is larger than 0.5 for certain frequencies. The uncertainty analysis has been performed using the Monte Carlo method of Supplement 1 to the ‘Guide to the Expression of Uncertainty in Measurement’ and correlations in the estimated impulse response have been determined. Our measurements extend previous work which deals with the characterization of 70 GHz oscilloscopes and the measurement of 100 GHz oscilloscopes up to the cut-off frequency of higher order modes.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.