In this paper the range of optimum lens shapes for theoretical pseudophakic eyes with high myopia is given as a function of intraocular lens (IOL) power using physiological and optical design criteria. Wave-front aberration in the plane of the exit pupil in pseudophakic eyes with high myopia is studied. Taking into account the defocus and spherical aberration coefficients, we have used a coefficient which gives us information about the image quality of optical systems. Considering an IOL diameter of 5$5 mm and excluding concave surfaces facing the retina, this coefficient and the anterior sagitta of each IOL are calculated as functions of the bending factor. Bending factors that give an anterior sagitta of the IOL lower than 0$5 mm and a coefficient lower than the coefficient of an emmetropic phakic theoretical eye are chosen as the optimum shapes of the IOL for each power.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.