Abstract

AbstractIn this paper, a quadruple‐potential waveform was investigated and optimized for the determination of gentamicin by reversed phase ion‐pair chromatography. Instead of a relatively high positive potential, a negative potential was adopted as a potential for the cleaning of gold working electrode. By this way, the formation of gold oxide resulting from the application of high positive potential during the analyte detection and electrode cleaning was greatly reduced, and therefore, the dissolution and recession of gold working electrode was also reduced. The good condition of gold working electrode achieved by this quadruple‐potential waveform can help us to obtain a good reproducibility. In order to acquire signal‐to‐noise ratio as high as possible, several waveform parameters affecting the detection of gentamicin were carefully selected. The analytical method has been applied to the determination of two real gentamicin samples, and good results with low relative standard deviation not more than 4% were obtained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.