Abstract
AbstractAn approach is described for optimizing chromatographic conditions for dual‐column analysis of U.S. EPA Method 608 pesticides. Sample throughput for environmental analysis of pesticides can be increased by performing the screening and confirmation analyses at the same time. This requires simultaneous injection, separation, and detection in a GC with dual inlets, columns, and detectors. Resolution maps for flow rate and temperature relationships of poorly resolved peak pairs for 5%‐phenyl methyl silicone and 7%‐cyanopropyl 7%‐phenyl methyl silicone capillary columns are presented. Resolution of target pesticides on the 5%‐phenyl methyl silicone column was much more sensitive to changes in flow and temperature‐ramp rates than was the 7%‐cyanopropyl 7%‐phenyl methyl silicone column. Optimal conditions resulted in a run time of approximately 20 min.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.