Abstract
ABSTRACTThis work demonstrated the improvement of the visualization of lesions by modulating the factors of an X-ray mammography imaging system using Taguchi analysis. Optimal combinations of X-ray operating factors in each group of level combination were determined using the Taguchi method, in which all factors were organized into only 18 groups, yielding analytical results with the same confidence as if each factor had been examined independently. The 4 considered operating factors of the X-ray machine were (1) anode material (target), (2) kVp, (3) mAs and (4) field of view (FOV). Each of these factors had 2 or 3 levels. Therefore, 54 (2×3×3×3 = 54) combinations were generated. The optimal settings were Rh as the target, 28 kVp, 80 mAs and 19×23 cm2 FOV. The grade of exposed mammographic phantom image increased from the automatic exposure control (AEC) setting 70.92 to 72.00 under the optimal setting, meeting the minimum standard (70.00) set by Taiwan's Department of Health. The average glandular dose (AGD) of the exposed phantom, 0.182 cGy, was lower than that, 0.203 cGy, under the AEC setting. The Taguchi method was extremely promising for the design of imaging protocols in clinical diagnosis.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.