Abstract
The performance and accuracy of quantum electronics is substantially degraded when the temperature of the electrons in the devices is too high. The electron temperature can be reduced with appropriate thermal anchoring and by filtering both the low frequency and radio frequency noise. Ultimately, for high performance filters the electron temperature can approach the phonon temperature (as measured by resistive thermometers) in a dilution refrigerator. In this application note, the method for measuring the electron temperature in a typical quantum electronics device using Coulomb blockade thermometry is described. This technique is applied to find the readily achievable electron temperature in the device when using the QFilter provided by QDevil. With our thermometry measurements, using a single GaAs/AlGaAs quantum dot in an optimized experimental setup, we determined an electron temperature of 28 ± 2 milli-Kelvin for a dilution refrigerator base temperature of 18 milli-Kelvin.
Published Version