Abstract
BackgroundEffective workflow management in a diagnostic pathology laboratory is critical to achieve rapid turnover while maintaining high quality. Fluorescence in situ hybridization analysis (FISH) is the preferred technique for detecting single chromosomal aberrations in diagnostic surgical pathology.Material and methodsFISH analysis applying a rapid hybridization protocol and using an automated whole-slide fluorescence scanning device (3DHISTECH, Sysmex, Switzerland) were implemented in our workflow. By analyzing 42 diagnostic cases, effects of two different scanning profiles on scanning time, and device memory usage were investigated. Manual signal counting (CaseViewer) and software based signal counting (FISHQuant) were compared.ResultsThe two scanning profiles, both including a Z-stack function, differed in their exposure time and digital gain. The “low profile” setting (LP) resulted in a significantly shorter scanning time and lower storage volume compared to the “high profile” (HP) setting, making the LP ideal for routine applications. Both signal counting methods (manual versus software based) provided similar cut-offs on a test-cohort of 13 samples.ConclusionScanning FISH slides provides good picture quality, reduces the analysis time and allows easy picture archiving and facilitates remote diagnostics, allowing an effective workflow.
Published Version (
Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have