Abstract
We propose an approximate approach to use redundant data outside the 1PI window within the exact Katsevich reconstruction framework. The proposed algorithm allows a flexible selection of the helical pitch, which is useful for clinical applications. Our idea is an extension of the one proposed by KOhler, Bontus, and Koken (2006). It is based on optimizing the contribution weights of convolution families used in exact Katsevich 3PI algorithms, so that the total weight of each Radon plane is as close to 1 as possible. Optimization is based on solving a least squares problem subject to linear constrains. Numerical evaluation shows good noise and artifact reduction properties of the proposed algorithm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.