Abstract

Instrumentation has been developed for the measurement of electron temperature and density profiles at the boundary of COMPASS-D tokamak plasmas from He I line intensities. The technique is applicable to measurements of ne and Te over the relevant ranges of 1017–1020 m−3 and 10–500 eV. Spectra from eight viewing chords viewing a thermal helium jet are recorded on a CCD sensor with ≈1 ms time resolution. A 0.6 m Czerny-Turner spectrometer, corrected for astigmatism using an internal, cylindrical lens optimizes throughput by availing of a 110×110 mm2, 300 l/mm grating. ne and Te profile measurements from L-mode (low) and quiescent H-mode (high) confinement plasmas are presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.