Abstract
AbstractThe use of sequentially coupled columns to achieve a binary liquid phase mixture has been simplified by the availability of zero dead volume fittings compatible with fused silica columns. The extent to which the velocity gradient through the coupled column affects the âapparentâ liquid phase ratio can be determined by graphic interpolation; the length ratio of the coupled column segments can then be adjusted so that the âapparentâ liquid phase ratio actually experienced by the solutes agrees with the targeted value. The fact that the direction of flow through the coupled column affects the chromatographic dispersion suggests that accepted generalizations on flow optimization throughout the column may not be precisely correct.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.