Abstract

Optimal conditions for low energy electron point source microscopy are investigated by the simulation and numerical reconstruction of holograms of phthalocyaninato polysiloxane, PcPS, a rod-like macromolecule. The effects of the electron energy, width of the electron beam and the detector size on the spatial resolution in the reconstructed images are modeled. We find that for electron energies around 200 eV, with the specimen 0.1 μm from the source, a screen recording the image in a cone of at least 15° half angle (7 cm lateral dimension at 10 cm from the source) with at least 512×512 pixel and 8-bit resolution will result in near atomic resolution.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.