Abstract

The Lawrence Livermore National Laboratory – Center for Accelerator Mass Spectrometry (LLNL/CAMS) 1 MV AMS system was converted from a biomedical AMS instrument to a natural abundance 14C spectrometer. The system is equipped with a gas-accepting hybrid ion source capable of measuring both solid (graphite) and gaseous (CO2) samples. Here we describe a series of experiments intended to establish and optimize 14CO2 measurement capabilities at natural abundance levels. A maximum instantaneous ionization efficiency of 8 % was achieved with 3 % CO2 in helium at a flow rate of approximately 220 µL/min (3.5 µg C/min). For modern materials (e.g., OX I) we measured an average of 240 ± 50 14C counts/µg C, equivalent to a total system efficiency of approximately 3 %. Experimental CO2 samples with F14C values ranging from 0.20 to 1.05 measured as both graphite and directly as CO2 gas produced equivalent values with an average offset of < 2σ.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.