Abstract

The Lawrence Livermore National Laboratory – Center for Accelerator Mass Spectrometry compact 1 MV biomedical accelerator mass spectrometer was repurposed and optimized for the semi-automated radiocarbon measurement of natural abundance environmental samples. Here we present results from 15 months of routine system operation and evaluate the system performance based on 30 sample wheels measured with consistent operating conditions over 7 months from August 2019 to March 2020. The average measurement precision was 2.7 ± 0.7‰. System accuracy was assessed with measurements of 14C-free materials, standard materials with 14C-content spanning 0.5 to 1.5 F14C, and by comparison to split samples measured with the 10 MV AMS system. Overall, this system can routinely measure samples as small as 300 µgC and less than 33000 years without the need for size-specific correction protocols.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.