Abstract

In the present work a simple model for calculation of the energy response of radiographic films was introduced. According to the model the energy response of a radiographic film is directly proportional to the optical density on the film and thus to the number of developed grains in the emulsion. The model was simulated by Monte Carlo method using MCNP code and the relative energy response of Kodak type 2 film under a few filters of A.E.R.E./R.P.S. film badge was calculated. The simulated responses were in agreement with the experimental data in the region of 30 keV–1.5 MeV. In the next stage a multi-element filter was simulated to optimize the energy response in the above energies. The energy response varied by 25% between 40 keV and 1.5 MeV. So the dose received by the film is equivalent to the desired true dose and there would be no need to the correction factors.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.