Abstract

In this study, we evaluated the performance of a digital chest imaging system using a contrast-detail (C-D) phantom. In the initial step, 76 sample images of the C-D phantom were produced by changing the doses from 0.5, 0.75, 1.0, 1.25, 1.5, to 2.0 times the dose for a screen-film (S/F) system. The sample images were analyzed by five radiological technologists and two medical physicists, and the image quality figure (IQF) was determined. The quality of each image was examined, and appropriate doses were determined from the calculated IQF to obtain the same image quality for other digital chest imaging systems. The method of determining IQF from C-D phantom analysis was very useful for comparing image quality and determining radiographic techniques.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.