Abstract

This study compares static conformal field (CF), intensity modulated radiotherapy (IMRT), and dynamic arcs (DA) for the stereotactic radiotherapy of base-of-skull meningiomas. Twenty-one cases of base-of-skull meningioma (median planning target volume [PTV] = 21.3 cm 3) previously treated with stereotactic radiotherapy were replanned with each technique. The plans were compared for Radiation Therapy Oncology Group conformity index (CI) and homogeneity index (HI), and doses to normal structures at 6 dose values from 50.4 Gy to 5.6 Gy. The mean CI was 1.75 (CF), 1.75 (DA), and 1.66 (IMRT) ( p < 0.05 when comparing IMRT to either CF or DA plans). The CI (IMRT) was inversely proportional to the size of the PTV (Spearman's rho = −0.53, p = 0.01) and at PTV sizes above 25 cm 3, the CI (IMRT) was always superior to CI (DA) and CI (CF). At PTV sizes below 25 cm 3, there was no significant difference in CI between each technique. There was no significant difference in HI between plans. The total volume of normal tissue receiving 50.4, 44.8, and 5.6 Gy was significantly lower when comparing IMRT to CF and DA plans ( p < 0.05). There was significantly improved dose sparing for the brain stem and ipsilateral temporal lobe with IMRT but no significant difference for the optic chiasm or pituitary gland. These results demonstrate that stereotactic IMRT should be considered to treat base-of-skull meningiomas with a PTV larger than 25 cm 3, due to improved conformity and normal tissue sparing, in particular for the brain stem and ipsilateral temporal lobe.